Gas Flow Calibrations
|
The PVTt systems
are best suited
for
calibrating critical flow venturis (CFV) or critical nozzles since
pressure at
the inlet of the standard varies between vacuum and 100 kPa as the
collection
tank fills. The Working Gas Flow Standard (WGFS) is normally used
for
other flow meter types, e.g. laminar flow elements, turbine meters, and
positive
displacement meters. >
The WGFS uses critical venturis and critical nozzles that have been calibrated in the PVTt standards as working standards or reference flow meters to calibrate customer flow meters. The uncertainty of the WGFS is 0.1% at flows less than 2000 L/min and 0.16% at flows larger than 2000 L/min. Pressure regulators provide CFV pressures up to 700 kPa. The gas source for the WGFS can be dried air from our compressor, nitrogen from a dewar (<150 L/min), or other gases from compressed gas cylinders (<100 L/min).
To read more about
the principles
of
operation and uncertainties of our gas flow standards and see sample
calibration reports see our Technical
Notes and Publications.
Turnaround time for a
single flow meter is
approximately 4
weeks: