NIST Sealed-Cell Fixed Point Development

Goals: Develop, maintain and disseminate ITS-90 primary and secondary fixed points using sealed-cell devices for triple point realizations.

Applications:

  • International Fixed Point Comparisons for gas-based triple points
  • Reference Material development
  • Fixed Point Dissemination to Secondary Standards Laboratories
  • Temperature Scale Research

Advantages:

  • Compact and robust fixed point devices
  • Easily Transportable
  • Minimum chance of contamination
  • Long term stability

Range: T=161.406 K to 13.8 K

NIST Sealed Triple Point Cells currently in place or under development:

Material
Triple Point
Status
Purity
Melting Range
e-H2 13.8033 K ITS-90 99.9999% -
e-D2* 18.689 K secondary 99.995% -
Ne 24.5561 K ITS-90 99.9995% 0.27 mK
O2 54.3584 K ITS-90 99.9999+% 0.06 mK
Ar 83.8058 K ITS-90 99.9999% 0.18 mK
Xe 161.406 K secondary 99.999% 0.5 mK

* In collaboration with the IMGC, Turin, Italy

Techniques:

  • Melting plateau
  • Pulsed heating
  • Adiabatic re-equilibration

Cell Designs:

  • AR and O2: All Stainless Steel, Types 316L and 304L
  • NE and H2: Type 316L SS with OFHC copper inserts, electron-beam welded (see Figure 1)
  • D2: Type 316L Vacuum-Arc-Remelt
  • Xe: All OFHC copper (NRC type)

Ne TP Cell and Wiring Platform

Figure 1. A Ne TP cell and wiring platform.

 

For more information, contact Weston Tew at weston.tew@nist.gov